• REPLACEMENT FOR

    • Analytical Industries [PSR-11-915]
    • Datascope [Multinex Analyzer]
    • GE (Datex-Ohmeda) [0237-2034-700 Excel Modulus,5250 RGM,5150 Handheld Monitor]
    • GE (Ohmeda) [5120, 5100, 5250 RGM, 4700 Oxicap, 7800, 5125, 5150, Handheld Monitor, 0237-2034-700, Excel Modulus]
    • Hewlett Packard [15201A]
    • Mercury Medical [10-103-07]
    • Pacifitech [PT-10]
    • Teledyne [R15 #C-41131]
    • Teleflex (Hudson) [5556]
    • VTI [100352, 101272, 100352]

    SPECS

    • Range of Measurement: 0 to 100%
    • Output Range: 14 to 20 mV
    • Zero Offset: ≤ 0.30 mV
    • Response Time: ≤ 17 seconds
    • Linearity: ≤ 3% full scale
    • Expected Lifetime: > 500,000 %O2 hours
    • Temperature Compensation: none
    • Operational Temperature Range: 10˚ to 40˚C
    • Storage Temperature Range: -20˚ to 50˚C
    • Humidity Range: 0% to 90% RH, non-condensing
    • Ambient Pressure Range: 700-1250 mbar
    • Interference: per EN ISO 80601-2-55
    • Electrical Interface: gold plated slip rings