REPLACEMENT FOR

  • Analytical Industries [PSR-11-915-1]
  • Drager [6803290, 6850645, Narkomed, Fabius GS, Tiro]
  • Hewlett Packard [15201A]
  • Invivo [Magnitude]
  • Mercury Medical [10-103-12]
  • Pacifitech [PT-24]
  • Teledyne [R31]
  • VTI [101372, 101372]

SPECS

  • Range of Measurement: 0 to 100% O2
  • Output Range: 12.2 to 18.1 mV
  • Zero Offset: ≤ 0.50 mV
  • Response Time: ≤ 15 seconds at 23±2°C
  • Linearity: ± 1.0%
  • Expected Lifetime: > 2,000,000 %O2 hours
  • Operational Temperature Range: 5° to 40°C
  • Storage Temperature Range: -15° to 50°C
  • Humidity Range: 5% to 95% RH, non-condensing
  • Interference: Per EN ISO 80601-2-55