REPLACEMENT FOR

  • Analytical Industries [PSR-11-915-4]
  • City Technology [MOX-7]
  • EnviteC [OOM-110]
  • GE (Ohmeda) [6050-0004-110, Modulus SE, 7900 Series, Aisys, Aestiva Series, SmartVent S-5, Aespire, Avance]
  • International Technologies [M-10]
  • Mercury Medical [10-103-10]
  • Pacifitech [PT-20A]
  • Teledyne [R30MED #B-72939]

SPECS

  • Range of Measurement: 0 to 100% O2
  • Output Range: 8.0 to 13.5mV
  • Zero Offset: > 0.15mV
  • Response Time: ≤ 15 s
  • Linearity: ± 2%
  • Expected Lifetime: > 900,000 %O2 hours
  • Operational Temperature Range: 10° to 40°C
  • Storage Temperature Range: -20° to 50°C
  • Humidity Range: 10% to 95% RH, noncondensing
  • Ambient Pressure Range: 666-1066 mbar
  • Interference: Per EN ISO 80601-2-55