REPLACEMENT FOR
- Figaro [KE-25]
SPECS
- Range of Measurement: 0 to 100% O2
- Output Range: 10.0 – 15.5 mV
- Zero Offset: < 0.5 mV
- Response Time: 14 ± 2 s
- Linearity: ± 2.0%
- Expected Lifetime: 5 years
- Operational Temperature Range: 5° – 40°C
- Storage Temperature Range: -15° to 50°C
- Humidity Range: 10 to 90% RH
- Ambient Pressure Range: 811-1216 hPa
- Interference: Per EN ISO 80601-2-55
- Electrical Interface: .062″ dual-pin Molex