REPLACEMENT FOR

  • Maxtec [MaxO2 ME, R140P02]

SPECS

  • Range of Measurement: 0 to 100%
  • Output Range: 10.5 to 15 mV
  • Zero Offset: ≤ 0.50 mV
  • Response Time: ≤ 15 s
  • Linearity: ≤ 2%
  • Expected Lifetime: 1,500,000 %O2 hours
  • Temperature Compensation: NTC
  • Operational Temperature Range: 15˚ to 40˚C
  • Storage Temperature Range: -15˚ to 50˚C
  • Humidity Range: 0 to 95% RH, non-condensing
  • Ambient Pressure Range: 800-1013 mbar
  • Interference: per EN ISO 80601-2-55
  • Electrical Interface: 3-pin stereo jack