Max-7
REPLACEMENT FOR
- Precision Medical [505327]
- Teledyne [R17MED #C-43690, AX/MX300 Series, R17MED #C-43690]
SPECS
- Range of Measurement: 0 to 100% O2
- Output Range: 8.0 to 12.0 mV
- Zero Offset: ≤ .2 mV
- Response Time: ≤ 12 s
- Linearity: ± 2.0%
- Expected Lifetime: < 6 years
- Temperature Compensation: NTC
- Operational Temperature Range: 0°C to 45°C
- Storage Temperature Range: -20°C to 50°C
- Humidity Range: 5% to 95% RH, non-condensing
- Interference: Per EN ISO 80601-2-55
- Electrical Interface: 3.5mm mono jack